STANDARD OPERATING PROCEDURE
Valid for:
2 years from
approval
Page 9 of 11
HE
Press SELECT key twice to display
force cell capacity of gauge
Does the displayed
force cell capacity
match with that of
force cell module
Apply test weight that is equal to
the displayed full-scale capacity weight
Used weight is lower than the full
capacity of the force cell module
Repeat auto-calibration with a
standard weight equal to the full
capacity of the force cell module
Used weight is higher than the full
capacity of the force cell module
Wait until device displays the applied
weight to indicate that calibration
process was successful
Repeatedly press SCROLL key until
correct force cell capacity is displayed