E. OPERATING INSTRUCTIONS
The following instructions cover the basic operation of the FDV
Digital Force Gage, allowing immediate use.
1. To turn the FDV on, press the POWER button. If there is no display or if
“LO BAT” is displayed, the battery may be low or not connected
securely.
CAUTION: When "HELP" is displayed it may indicate either:
1) Force Cell Module is not securely connected.
2) Force Cell Module has been overloaded and is damaged.
Call Wagner Instruments for instructions.
CAUTION: "LoAd" will appear on the display when the gage is first
activated if the gage has been disconnected from the battery or AC
power for an extended period of time, and the set-up menu has
reverted to default settings. User must reset "LoAd" (see Section F.)
and should check other menu settings before testing to assure that
previously selected values are as desired. The FDV cannot be shut
off in this mode until a value for "LoAd" has been selected.
RECOMMENDATION: Warm-up time for the FDV is recommended.
After initial power-on, allow 2 - 3 minutes for electronics to stabilize.
This will improve accuracy of readings.
RECOMMENDATION: Exercising the FDV is recommended. After
initial warm up apply approximately 100% of capacity load prior to
use. This eliminates residual hysteresis in the Force Cell Module.
2. To turn the FDV off, press the POWER button. The FDV will shut down
if set to do so and if no buttons have been pressed for the selected
elapsed time. (See Section F. 4.3) All user defined functions remain
stored in memory while the gage is shut off.
3. With the FDV oriented in the same position as it will be during use,
press the ZERO button. This automatically tares the weight of any
attachment on the shaft, sets the display to zero, and clears all peak
readings.