The RS232 data is transferred in the following format:
-XXXXXX YY Peak <CR> <LF>
Line Feed
Carriage return
“Peak” will be transmitted
if in a peak mode
Units of measure
Force value (5 digits and decimal point)
or “HELP” if the gage is overloaded
Space for a compression force
minus sign for a tension force
J.2 Mitutoyo Output
The Mitutoyo data format is a modified BCD (binary coded decimal)
format that is used in communication with Mitutoyo statistical process
control printers.
To transmit the value shown on the FDV display to a Mitutoyo printer,
the user may either:
A. Press the PEAK and ZERO buttons simultaneously,
or
B. Press the DATA button on the Mitutoyo processor.
If the gage is overloaded there will be no response.